Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-222 DMM7510-901-01 Rev. B / May 2015
localnode.linefreq
This attribute contains the power line frequency setting that is used for NPLC calculations.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Power cycle Not applicable Not applicable
Usage
frequency = localnode.linefreq
frequency
The detected line frequency: 50 or 60
Details
The instrument automatically detects the power line frequency (either 50 Hz or 60 Hz) when the
instrument is powered on.
Example
frequency = localnode.linefreq
print(frequency)
Reads line frequency setting.
Also see
None
localnode.model
This attribute stores the model number.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
model = localnode.model
model
The model number of the instrument
Example
print(localnode.model)
Outputs the model number of the local node. For example:
DMM7510
Also see
localnode.serialno (on page 8-225)