EasyManuals Logo

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #723 background imageLoading...
Page #723 background image
Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-110 DMM7510-901-01 Rev. B / May 2015
dmm.digitize.readwithtime()
This function initiates digitize measurements and returns the last actual measurement and time information in
UTC format without using the trigger mode.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
reading, seconds, fractional = dmm.digitize.readwithtime()
dmm.digitize.readwithtime(bufferName)
reading
The last reading of the measurement process
seconds
Seconds in UTC format
fractional
Fractional seconds
bufferName
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer; if no buffer is specified, this parameter
defaults to
defbuffer1
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command initiates digitize measurements using the present function setting, stores the readings
in a reading buffer, and returns the last reading.
The dmm.digitize.count attribute determines how many measurements are performed.
When you use a reading buffer with a command or action that makes multiple readings, all readings
are available in the reading buffer. However, only the last reading is returned as a reading with the
command.
If you define a specific reading buffer, the reading buffer must exist before you make the
measurement.
Example
print(dmm.digitize.readwithtime())
Print the last digitize measurement and time information in
UTC format, which will look similar to:
-0.0003882925875 1415795836 0.946164546
Also see
dmm.digitize.count (on page 8-84)
dmm.measure.readwithtime() (on page 8-178)
trigger.model.load() — SimpleLoop (on page 8-299)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB