Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 2:
General operation
DMM7510-901-01 Rev. B / May 2015 2-133
Display results of two measure functions
The Model DMM7510 allows you to make and display two measurements from different functions.
The measurements are displayed on the front panel and stored in the reading buffers.
To access the dual measurement capability, swipe the lower half of the Home screen to the
SECONDARY swipe screen. This feature is only available from the front panel of the instrument.
Depending on the selected functions, a relay may click when the instrument switches between the
measurement types. Leaving secondary measurements on for extended periods may shorten the life
of the relays.
Figure 98: SECONDARY swipe screen