6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-178 DMM7510-901-01 Rev. B / May 2015
:TRIGger:BLENder<n>:CLEar
This command clears the blender event detector and resets the overrun indicator of blender <n>.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:BLENder<n>:CLEar
The blender number (1 or 2)
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
Clears the event detector for blender 2.
Also see
None
:TRIGger:BLENder<n>:MODE
This command selects whether the blender performs OR operations or AND operations.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Usage
:TRIGger:BLENder<n>:MODE <operation>
:TRIGger:BLENder<n>:MODE?
The blender number (1 or 2)
• OR
•