Page #401 background image
Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-74 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:BIAS:LEVel
This command selects the amount of current the instrument sources when it makes measurements.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
10 µA
Usage
[:SENSe[1]]:<function>:BIAS:LEVel <n>
[:SENSe[1]]:<function>:BIAS:LEVel DEFault
[:SENSe[1]]:<function>:BIAS:LEVel MINimum
[:SENSe[1]]:<function>:BIAS:LEVel MAXimum
[:SENSe[1]]:<function>:BIAS:LEVel?
[:SENSe[1]]:<function>:BIAS:LEVel? DEFault
[:SENSe[1]]:<function>:BIAS:LEVel? MINimum
[:SENSe[1]]:<function>:BIAS:LEVel? MAXimum
<function>
The function to which the setting applies; see Functions
<n>
The bias level: 10 µA, 100 µA, or 1 mA
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
Selects the amount of current that is sourced by the instrument to make measurements.
Example
DIOD:BIAS:LEVel 0.0001
For the diode functions, sets a bias level of
100
µ
A.
Also see
[:SENSe[1]]:<function>:BIAS:ACTual? (on page 6-73)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals