Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 3:
Functions and features
DMM7510-901-01 Rev. B / May 2015 3-63
Switching between measurement methods
The measurement methods that are available to you depend on how you are controlling the
instrument.
If you are using the front panel to control the instrument, you can choose any of the measurement
methods.
If you are using a remote interface to control the instrument, you can only use the trigger model
measurement method. When you switch to a remote interface, the trigger model measurement
method is automatically selected. If you switch from remote control to front-panel control, the trigger
model measurement method remains selected.
If you are running a script, the instrument automatically switches to the trigger model measurement
method.
Using the front panel:
1. Press the front-panel TRIGGER key for 2 s. A dialog box displays the available trigger methods.
The presently selected method is highlighted.
2. Select the method you want to use.
3. If the instrument is in remote control, the instrument displays a confirmation dialog box. Select
Yes to change to local control.
Triggering
Triggering allows you to start and synchronize measure actions on one or more instruments with a
trigger event or a combination of trigger events that you set. This section describes some of the
options available for triggering, including command interface triggering, timers, analog trigger, and
event blenders.
Command interface triggering
A command interface trigger event occurs when:
A GPIB GET command is detected (GPIB only)
A VXI-11 device_trigger method is invoked (VXI-11 only)
A *TRG message is received
To use a command interface trigger event as an input stimulus for another trigger object, set the
stimulus as TSP event trigger.EVENT_COMMAND or the SCPI event COMMAND. To ensure that
trigger commands that are issued over the command interface are processed in the correct order, the
instrument does not generate a trigger event until:
The trigger command is executed
TSP only: trigger.wait() retrieves the trigger command from the command queue before it
would normally be executed

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB