7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-187
:TRIGger:BLOCk:BRANch:LIMit:CONStant
This command defines a trigger model block that goes to a specified block if a measurement meets preset criteria.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:LIMit:CONStant <blockNumber>, <limitType>, <limitA>,
<limitB>, <branchToBlock>
:TRIGger:BLOCk:BRANch:LIMit:CONStant <blockNumber>, <limitType>, <limitA>,
<limitB>, <branchToBlock>, <measureBlock>
<blockNumber>
The sequence of the block in the trigger model
<limitType>
The type of limit (ABOVe, BELow, INside, or OUTside)
<limitA>
The limit that the measurement is tested against; if limitType is set to:
ABOVe: This value is ignored
BELow: The measurement must be below this value
INside: The low limit that the measurement is compared against
OUTside: The low limit that the measurement is compared against
<limitB>
The upper limit that the measurement is tested against; if limitType is set to:
ABOVe: The measurement must be above this value
BELow: This value is ignored
INside: The high limit that the measurement is compared against
OUTside
: The high limit that the measurement is compared against
<branchToBlock>
The block number of the trigger model block to execute when the measurement
meets the defined criteria
<measureBlock>
The block number of the measure or digitize block that makes the measurements to
be compared; if this is 0 or undefined, the trigger model uses the previous measure
or digitize block
Details
The branch-on-constant-limits block goes to a branching block if a measurement meets the criteria
set by this command.
The type of limit can be:
Above: The measurement is above the value set by limit B; limit A must be set, but is ignored
when this type is selected
Below: The measurement is below the value set by limit A; limit B must be set, but is ignored
when this type is selected