8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-328 DMM7510-901-01 Rev. B / May 2015
trigger.timer[N].clear()
This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.timer[N].clear()
Trigger timer number (1 to 4)
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
Also see
trigger.timer[N].count (on page 8-328)
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event or is enabled as
a timer or alarm.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
Number of times to repeat the trigger (0 to 1,048,575)
Trigger timer number (1 to 4)
Details
If count is set to a number greater than 1, the timer automatically starts the next trigger timer delay
at the expiration of the previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.