Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-289
trigger.model.load() — Empty
This function clears the trigger model.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.model.load("Empty")
Details
When you load this predefined trigger model, any blocks that have been defined in the trigger model
are cleared so the trigger model has no blocks defined.
Example
trigger.model.load("Empty")
print(trigger.model.getblocklist())
Clear the trigger model to have no blocks defined.
Output: EMPTY
Also see
None