MM7510 7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-237
:TRIGger:TIMer<n>:DELay
This command sets and reads the timer delay.
Type Affected by Where saved Default value
Instrument reset
Usage
:TRIGger:TIMer<n>:DELay <interval>
:TRIGger:TIMer<n>:DELay?
Trigger timer number (1 to 4)
Delay interval (8e-6 s to 100,000 s)
Details
Each time the timer is triggered, it uses this delay period.
Reading this command returns the delay interval that will be used the next time the timer is triggered.
Example
Set trigger timer 2 to delay for 50 µs.
Also see
None
:TRIGger:TIMer<n>:STARt:FRACtional
This command configures an alarm or a time in the future when the timer will start.
Type Affected by Where saved Default value
Instrument reset
Usage
:TRIGger:TIMer<n>:STARt:FRACtional <time>
:TRIGger:TIMer<n>:STARt:FRACtional?
Trigger timer number (1 to 4)
The time in fractional seconds (0 to < 1 s)