dmm.measure.limit[Y].clear()
This function clears the results of the limit test defined by Y.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
dmm.measure.limit[Y].clear()
Y
Limit number: 1 or 2
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
Use this command to clear the test results of limit Y when the limit auto clear option is turned off. Both
the high and low test results are cleared.
To avoid the need to manually clear the test results for a limit, turn the auto clear option on.
Example
See dmm.measure.limit[Y].low.value (on page 8-164) for an example of how to use this command.
Also see
dmm.digitize.limit[Y].clear() (on page 8-94)
dmm.measure.limit[Y].autoclear (on page 8-159)