Model D
MM7510 7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-51
ROUTe subsystem
The ROUTe subsystem selects which set of input and output terminals to enable (front panel or rear
panel).
:ROUTe:TERMinals
This command describes which set of input and output terminals the instrument is using.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:ROUTe:TERMinals?
Details
You must use the front-panel TERMINALS button to change which set of terminals the instrument
reads.
This query returns the set of input and output terminals that the instrument is using. If the instrument
is using the front-panel terminals, the return is:
FRON
If the instrument is using the rear-panel terminals, the return is:
REAR
Example
:ROUT:TERM?
Query to verify which terminals are used.
Output if the rear terminals are used:
REAR
Also see
None

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals