6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-184 DMM7510-901-01 Rev. B / May 2015
Details
When the trigger model reaches the Counter Reset block, it resets the count of the specified Branch
on Counter block to zero.
Example
TRIG:LOAD "EMPTY"
TRIG:BLOC:BUFF:CLEAR 1
TRIG:BLOC:MEAS 2
TRIG:BLOC:BRAN:COUN 3, 5, 2
TRIG:BLOC:DEL:CONS 4, 1
TRIG:BLOC:BRAN:COUN 5, 3, 2
TRIG:BLOC:BRAN:COUN:RES 6, 3
Reset trigger model settings.
Clear defbuffer1 at the beginning of the
trigger model.
Loop and take 5 readings.
Delay a second.
Loop three more times back to block 2.
Reset block 3 to 0.
Also see
:TRIGger:BLOCk:BRANch:COUNter (on page 6-182)
:TRIGger:BLOCk:BRANch:COUNter:COUNt? (on page 6-183)
:TRIGger:BLOCk:BRANch:DELTa
This command defines a trigger model block that goes to a specified block if the difference of two measurements
meets preset criteria.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:DELTa <blockNumber>, <targetDifference>, <branchToBlock>
:TRIGger:BLOCk:BRANch:DELTa <blockNumber>, <targetDifference>, <branchToBlock>,
<measureBlock>
<blockNumber>
The sequence of the block in the trigger model
<targetDifference>
The value against which the block compares the difference between the
measurements
<branchToBlock>
The block number of the trigger model block to execute when the difference
between the measurements is less than or equal to the
<targetDifference>
<measureBlock>
The block number of the measure or digitize block that makes the
measurements to be compared; if this is 0 or undefined, the trigger model uses
the previous measure or digitize block