EasyManuals Logo

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #872 background imageLoading...
Page #872 background image
Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-259
trigger.clear()
This function clears any pending command triggers.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.clear()
Details
A command trigger indicates if a trigger event has been detected over a command interface since the
last trigger.wait() command was sent. Command triggers are generated by:
Sending *TRG over a remote interface
GET bus commands
VXI-11 device trigger commands
trigger.clear() clears the command triggers and discards the history of trigger events.
Example
*TRG
print(trigger.wait(1))
trigger.clear()
print(trigger.wait(1))
Generate a trigger event.
Check if there are any pending trigger
events.
Output: true
Clear any pending command triggers.
Check if there are any pending trigger
events.
Output:
false
Also see
trigger.wait() (on page 8-345)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals