8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter Referenc
e Manual
8-338 DMM7510-901-01 Rev. B / May 2015
trigger.timer[N].wait()
This function waits for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
triggered = trigger.timer[N].wait(timeout)
triggered
Trigger detection indication
N
Trigger timer number (1 to 4)
timeout
Maximum amount of time in seconds to wait for the trigger
Details
If one or more trigger events were detected since the last time trigger.timer[N].wait() or
trigger.timer[N].clear() was called, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed.
This is true regardless of the number of events detected.
Example
triggered = trigger.timer[3].wait(10)
print(triggered)
Waits up to 10 s for a trigger on timer 3.
If false is returned, no trigger was detected
during the 10 s timeout.
If
true
is returned, a trigger was detected.
Also see
trigger.timer[N].clear() (on page 8-328)
trigger.tsplinkin[N].clear()
This function clears the event detector for a LAN trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.tsplinkin[N].clear()
N
The trigger line (1 to 3) to clear