8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-262 DMM7510-901-01 Rev. B / May 2015
Details
If this is true, an event was ignored because the event detector was already in the detected state
when the event occurred.
This is an indication of the state of the event detector built into the line itself. It does not indicate if an
overrun occurred in any other part of the trigger model or in any other detector that is monitoring the
event.
Example
overrun = trigger.digin[1].overrun
print(overrun)
If there is no trigger overrun on digital
input 1, the output is:
false
Also see
digio.line[N].mode (on page 8-52)
digio.line[N].reset() (on page 8-54)
Digital I/O port configuration (on page 3-49)
trigger.digin[N].clear() (on page 8-260)
trigger.digin[N].wait()
This function waits for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
triggered = trigger.digin[N].wait(timeout)
triggered
Trigger detected: true
No triggers detected during the timeout period:
false
N
Digital I/O trigger line (1 to 6)
timeout
Timeout in seconds
Details
This function pauses for up to timeout seconds for an input trigger. If one or more trigger events are
detected since the last time trigger.digin[N].wait() or trigger.digin[N].clear() was
called, this function returns a value immediately. After waiting for a trigger with this function, the event
detector is automatically reset and is ready to detect the next trigger. This is true regardless of the
number of events detected.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB