Section
3: Functions and features Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
3-46 DMM7510-901-01 Rev. B / May 2015
From the front panel:
1. Press the MENU key.
2. Under System, select Calibration.
3. Select Scheduling Action. To have the instrument:
ï‚§ Prompt you to run auto calibration: Select Notify.
ï‚§ Run auto calibration at a specific time: Select Run.
ï‚§ To stop scheduling: Select None. If you select None, you do not need to make additional
settings.
4. Select Scheduling Interval.
5. Select the interval.
6. Select Scheduled Time to select the time when the auto calibration will run or when you will be
prompted to run it.
To review the next schedule time and date, see the information listed next to Next Run.
Using SCPI commands:
Refer to :ACAL:SCHedule (on page 6-21).
Using TSP commands:
Refer to acal.schedule() (on page 8-13).
Reviewing calibration information
The Calibration screen displays information about the last autocalibration and factory calibrations that
were run and the present status. For detail on this screen, refer to System Calibration menu
(on page
2-53).
For autocalibration, you can also access this information from the commands in the SCPI
ACAL
subsystem (on page 6-15) or the TSP acal.* commands.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals