Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 3: Fun
ctions and features
DMM7510-901-01 Rev. B / May 2015 3-43
Deleting a configuration index
Use the following procedure to delete a configuration index from MyMeasList.
Using the front panel to delete a configuration index:
1. Press the MENU key.
Under Measure, select Config Lists. The MEASURE CONFIGURATION LISTS screen is
displayed.
2. Choose Select List. A menu of available configuration lists is displayed.
3. Select MyMeasList. The configuration indexes are displayed.
4. Select the index to be deleted.
5. Select Delete Index.
Using remote commands for configuration list operations
You can use the following remote commands to create and maintain configuration lists.
Action SCPI command
TSP command
Create a configuration list
[:SENSe[1]]:CONFiguration:LIST:CREate (on page 6-116)
dmm.measure.configlist.create() (on page 8-139)
Restore the settings in a configuration list to the
instrument
[:SENSe[1]]:CONFiguration:LIST:RECall (on page 6-118)
dmm.measure.configlist.recall() (on page 8-142)
View the contents of a configuration list index as
TSP commands
[:SENSe[1]]:CONFiguration:LIST:QUERy? (on page 6-117)
dmm.measure.configlist.query() (on page 8-141)
Delete a configuration list or an index in a
configuration list
[:SENSe[1]]:CONFiguration:LIST:DELete (on page 6-116)
dmm.measure.configlist.delete() (on page 8-140)
View available configuration lists [:SENSe[1]]:CONFiguration:LIST:CATalog? (on page 6-115)
dmm.measure.configlist.catalog() (on page 8-138)
Determine the number of configuration indexes
in a configuration list
[:SENSe[1]]:CONFiguration:LIST:SIZE? (on page 6-119)
dmm.measure.configlist.size() (on page 8-143)
Save a configuration list
[:SENSe[1]]:CONFiguration:LIST:STORe (on page 6-120)
dmm.measure.configlist.store() (on page 8-143)
Saving a configuration list
Configuration lists are lost when you turn the instrument off and turn it on again. Save a configuration
list by creating a configuration script (with TSP command createconfigscript() or front panel)
or using the *SAV and *RCL commands (SCPI). A configuration script saves the settings of the
instrument, including all configuration lists. See Saving setups (on page 2-150
) for additional
information.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB