8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
8-154 DMM7510-901-01 Rev. B / May 2015
dmm.measure.fourrtd
This attribute defines the type of 4-wire RTD that is being used
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
dmm.RTD_PT100
Usage
RTDType = dmm.measure.fourrtd
dmm.measure.fourrtd = RTDType
• PT100: dmm.RTD_PT100
• PT385: dmm.RTD_PT385
• PT3916: dmm.RTD_PT3916
• D100: dmm.RTD_D100
• F100: dmm.RTD_F100
•
User-specified type: dmm.RTD_USER
Functions
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_DIGITIZE_VOLTAGE
Details
The transducer type must be set to temperature and the transducer must be set to 4-wire RTD before
you can set the RTD type.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_FOURRTD
dmm.measure.fourrtd = dmm.RTD_PT3916
Set the measure function to temperature.
Set the transducer type to 4-wire RTD.
Set the RTD type to PT3916.
Also see
dmm.measure.rtdalpha (on page 8-184)
dmm.measure.rtdbeta (on page 8-185)
dmm.measure.rtddelta (on page 8-186)
dmm.measure.rtdzero (on page 8-187)
dmm.measure.transducer (on page 8-197)