7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
DMM7510-901-01 Rev. B / May 2015 6-103
[:SENSe[1]]:<function>:SRATe
This command defines the precise acquisition rate at which the digitizing measurements are made.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1,000,000
Usage
[:SENSe[1]]:<function>:SRATe <n>
[:SENSe[1]]:<function>:SRATe?
[:SENSe[1]]:<function>:SRATe? DEFault
[:SENSe[1]]:<function>:SRATe? MINimum
[:SENSe[1]]:<function>:SRATe? MAXimum
The function to which the setting applies; see Functions
1,000 to 1,000,000 readings per second
Functions
Details
The sample rate determines how fast the Model DMM7510 acquires a digitized reading.
Set the sample rate before setting the aperture. If the aperture setting is too high for the selected
sample rate, it is automatically adjusted to the highest aperture that can be used with the sample rate.
Example
DIG:CURR:SRATE 1000000
DIG:CURR:APER AUTO
DIG:COUN 10
Set the digitize function to measure current.
Set the sample rate to 1,000,000, with a
count of 10, and automatic aperture.
Make a digitize measurement.
Also see
[:SENSe[1]]:<function>:APERture (on page 6-53)