Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-113
[:SENSe[1]]:<function>:UNIT
This command sets the units of measurement that are displayed on the front panel of the instrument and stored in
the reading buffer.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Voltage: VOLT
Temperature: CELSius
Usage
[:SENSe[1]]:<function>:UNIT <unitOfMeasure>
[:SENSe[1]]:<function>:UNIT?
<function>
The function to which the setting applies; see Functions
<unitOfMeasure>
Temperature: KELVin, CELSius, or FAHRenheit
Digitize voltage, AC voltage, and DC voltage:
VOLT
or
DB
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
The change in measurement units is displayed when the next measurement occurs. You can only
change the units for the voltage, temperature, and digitize voltage functions. Other functions have a
fixed units setting that cannot be changed.
Example
VOLT:UNIT DB
Changes the front-panel display and buffer
readings for DC voltage measurements to be
displayed in decibels.
Also see
None

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals