EasyManua.ls Logo

Keithley DMM7510

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-144 DMM7510-901-01 Rev. B / May 2015
:SYSTem:EVENtlog:SAVE
This command saves the event log to a file on a USB flash drive.
Type Affected by Where saved Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:EVENtlog:SAVE "<filename>"
:SYSTem:EVENtlog:SAVE "<filename>", <eventType>
<filename>
A string that holds the name of the file to be saved
<eventType>
Limits the event log entries that are saved to specific types; set to:
ERRor: Saves only error entries
WARNing: Saves only warning entries
INFormational: Saves only informational entries
ALL: Saves all event log entries (default)
Details
This command saves all event log entries to a USB flash drive.
If you do not define an event type, the instrument saves all event log entries.
The extension .csv is automatically added to the file name.
Example
SYST:EVEN:SAVE "/usb1/July_error_log", ERR
Saves the error events in the event log to a
file on the USB flash drive named
July_error_log.csv.
Also see
:SYSTem:CLEar (on page 6-136)