7½ Digit Graphical Sampling Multimeter Reference Manual Section 6: SCPI comm
DMM7510-901-01 Rev. B / May 2015 6-33
:CALCulate[1]:<function>:MATH:MBFactor
This command specifies the offset, b, for the y = mx + b operation.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Usage
:CALCulate[1]:<function>:MATH:MBFactor <n>
:CALCulate[1]:<function>:MATH:MBFactor DEFault
:CALCulate[1]:<function>:MATH:MBFactor MINimum
:CALCulate[1]:<function>:MATH:MBFactor MAXimum
:CALCulate[1]:<function>:MATH:MBFactor?
:CALCulate[1]:<function>:MATH:MBFactor? DEFault
:CALCulate[1]:<function>:MATH:MBFactor? MINimum
:CALCulate[1]:<function>:MATH:MBFactor? MAXimum
The function to which the setting applies; see Functions
The offset for the y = mx + b operation; the valid range is −1e12 to +1e12
Functions
Details
This attribute specifies the offset (b) for an mx + b operation.
The mx + b math operation lets you manipulate normal display readings (x) mathematically according
to the following calculation:
y = mx + b
Where:
• y is the displayed result
• m is a user-defined constant for the scale factor
• x is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
• b is the user-defined constant for the offset factor