8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-318 DMM7510-901-01 Rev. B / May 2015
trigger.model.setblock() — trigger.BLOCK_DIGITIZE
This function defines a trigger block that makes a measurement using a digitize function.
Type TSP-Link accessible Affected by Where saved Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_DIGITIZE)
trigger.model.setblock(blockNumber, trigger.BLOCK_DIGITIZE, bufferName)
trigger.model.setblock(blockNumber, trigger.BLOCK_DIGITIZE, bufferName, count)
blockNumber
The sequence of the block in the trigger model
bufferName
The name of the buffer, which must be an existing buffer; if no buffer is defined,
defbuffer1
is used
count
The number of digitize readings to make before moving to the next block in the
trigger model; set to a specific value or infinite (trigger.COUNT_INFINITE) or
stop infinite (
trigger.COUNT_STOP
)
Details
When trigger model execution reaches the block:
1. The instrument begins making a measurement.
2. The trigger model execution waits for the measurement to complete.
3. The instrument places the measurement into the specified reading buffer, which cannot be of the
writable buffer style.
If you are defining a user-defined reading buffer, you must create it before you define this block.
When you set the count to a finite value, trigger model execution remains at the block until all
measurements are complete. If you set the count to infinite, the trigger model executes subsequent
blocks and measurements continue in the background until the trigger model execution reaches
another digitize block or until the trigger model ends.
A digitize function (digitize voltage or digitize current) must be selected before you run a trigger model
that contains this block. You cannot have a measure block and a digitize block in the same trigger
model.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB