Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 2:
General operation
DMM7510-901-01 Rev. B / May 2015 2-89
Modifying, repairing, or removing Keithley I/O Layer software
The Keithley I/O Layer interconnects many other installers.
To remove all the KIOL components, you need to uninstall the following applications using Control
Panel:
National Instruments NI
TM
IVI Compliance Package
National Instruments NI-VISA
TM
Run-Time Engine
IVI Shared Components
Visa Shared Components
Keithley SCPI Driver
After uninstalling components, reboot the computer.
Determining the command set you will use
You can change the command set that you use with the Model DMM7510. The remote command sets
that are available include:
SCPI: An instrument-specific language built on the SCPI standard.
TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
You cannot combine the command sets.
As delivered, the Model DMM7510 is set to work with the Model DMM7510 SCPI command set.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select the button next to Command Set.
4. Select the command set.
5. You are prompted to reboot.
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.