When making measurements in high-energy circuits, use test leads that meet the following
requirements:
Test leads should be fully insulated.
Only use test leads that can be connected to the circuit (for example, alligator clips and spade
lugs) for hands-off measurements.
Do not use test leads that decrease voltage spacing. These diminish arc protection and create a
hazardous condition.
Power circuit test procedure
Use the following procedure when testing power circuits:
1. Turn off power to the circuit using the regular installed connect-disconnect device. For example,
remove the device's power cord or turn off the power switch.
2. Attach the test leads to the circuit under test. Use appropriate safety-rated test leads for this
application. If over 42 V, use double-insulated test leads or add an additional insulation barrier for
the operator.
3. Set the Model DMM7510 to the proper function and range.
4. Power the circuit using the installed connect-disconnect device, and make measurements without
disconnecting the multimeter.
5. Remove power from the circuit using the installed connect-disconnect device.
6. Disconnect the test leads from the circuit under test.
DC voltage measurements
This section describes how you can set up DC voltage measurements from the front panel.
Inputs: Do not apply more than 1000 V
peak
between INPUT HI and LO. Failure to observe this
caution may result in instrument damage.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB