Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-364 DMM7510-901-01 Rev. B / May 2015
tspnet.write()
This function writes a string to the remote instrument.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
tspnet.write(connectionID, inputString)
connectionID
The connection ID returned from
tspnet.connect()
inputString
The string to be written
Details
The tspnet.write() function sends inputString to the remote instrument. It does not wait for
command completion on the remote instrument.
The Model DMM7510 sends inputString to the remote instrument exactly as indicated. The
inputString must contain any necessary new lines, termination, or other syntax elements needed
to complete properly.
Because tspnet.write() does not process output from the remote instrument, do not send
commands that generate too much output without processing the output. This command can stop
executing if there is too much unprocessed output from previous commands.
Example
tspnet.write(myID, "runscript()\r\n")
Commands the remote instrument to execute
a command or script named runscript() on
a remote device identified in the system as
myID
.
Also see
tspnet.connect() (on page 8-354)
tspnet.read() (on page 8-357)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals