6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-16 DMM7510-901-01 Rev. B / May 2015
:ACAL:LASTrun:TEMPerature:INTernal?
This command returns the internal temperature of the instrument when auto calibration was run.
Type Affected by Where saved Default value
Query only Not applicable Nonvolatile memory Not applicable
Usage
:ACAL:LASTrun:TEMPerature:INTernal?
Details
The temperature is displayed in Celsius (°C).
The instrument updates the internal temperature value when the instrument refreshes autozero. If
autozero is set to off or if autozero is not available for the selected function (such as capacitance,
continuity, frequency or period), the internal temperature value is not updated.
Example
ACAL:LAST:TEMP:INT?
Returns the internal temperature of the
instrument when auto calibration was last run.
Example output:
63.167084
Also see
:ACAL:RUN (on page 6-20)
Auto calibration (on page 3-44)