6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter Refe
6-234 DMM7510-901-01 Rev. B / May 2015
:TRIGger:STATe?
This command returns the present state of the trigger model.
Type Affected by Where saved Default value
Usage
:TRIGger:STATe?
Details
This command returns the state of the trigger model. The instrument checks the state of a started
trigger model every 100 ms.
This command returns the trigger state and the block that the trigger model last executed.
The trigger model states are:
• Idle: The trigger model is stopped.
• Running: The trigger model is running.
• Waiting: The trigger model has been in the same wait block for more than 100 ms.
• Empty: The trigger model is selected, but no blocks are defined.
• Building: Blocks have been added.
• Failed: The trigger model is stopped because of an error.
• Aborting: The trigger model is stopping because of a user request.
• Aborted: The trigger model is stopped because of a user request.
Example
An example output if the trigger model is inactive and ended at block 9 is:
IDLE;IDLE;9
Also see
None
:TRIGger:TIMer<n>:CLEar
This command clears the timer event detector and overrun indicator for the specified trigger timer number.
Type Affected by Where saved Default value
Usage
:TRIGger:TIMer<n>:CLEar
Trigger timer number (1 to 4)