7½ Digit Graphical Sampling Multimeter Reference Manual Section 9:
Frequently asked questions (FAQs)
DMM7510-901-01 Rev. B / May 2015 9-5
What to do if the GPIB controller is not recognized?
If the hardware is not recognized by the computer:
1. Uninstall the software drivers.
2. Reboot the computer.
3. Check for newer drivers on the vendor’s website. Check that the drivers are valid for the
operating system you have and any updates that might be necessary. This information is typically
found in the readme file that comes with the drivers.
4. Install software drivers.
5. Reboot the computer.
6. Plug in the hardware.
If it is still not recognized, you can try a different computer using a different operating system to rule
out operating system issues.
If this does not resolve the issue, contact the vendor of the GPIB controller for assistance.
I'm receiving GPIB timeout errors. What should I do?
If your GPIB controller is recognized by the operating system, but you get a timeout error when you
try to communicate with the instrument, check the following:
1. Confirm that the GPIB address you assigned to the instrument is unique and between the range
of 0 to 30. It should not be 0 or 21 because they are common controller addresses.
2. Check cabling connection. GPIB cables are heavy and can fall out of the connectors if they are
not screwed in securely.
3. Substitute cables to verify cable integrity. For example, if you can send and receive ASCII text,
but you cannot do a binary transfer, check your program and the decoding of the binary data. If
that does not resolve the problem, try another cable. ASCII text only uses seven data lines in the
cable; the binary transfer requires all eight lines.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB