EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Reference Manual

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #475 background imageLoading...
Page #475 background image
Section
6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter Referenc
e Manual
6-148 DMM7510-901-01 Rev. B / May 2015
:SYSTem:POSetup
This command selects the defaults that are used when you power on the instrument.
Type Affected by Where saved Default value
Command and query
Not applicable
Nonvolatile memory
RST
Usage
:SYSTem:POSetup <name>
:SYSTem:POSetup?
<name>
Which setup to restore when you power on the instrument:
• Power on to *RST defaults: RST
• Stored setup 0: SAV0
• Stored setup 1: SAV1
• Stored setup 2: SAV2
• Stored setup 3: SAV3
• Stored setup 4:
SAV4
Details
When you select RST, the instrument restores settings to their default values when the instrument is
powered on.
When you select a SAV option, the settings in the selected saved setup are applied when the
instrument is powered on. The settings are saved using the *SAV command.
Example
SYST:POS SAV1
Set the instrument to restore the settings
that are saved in the stored setup 1 when
the instrument is powered on.
Also see
*SAV (on page 6-14)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals