7½ Digit Graphical Sampling Multimeter Reference Manual Section 2:
DMM7510-901-01 Rev. B / May 2015 2-119
Figure 83: Rear panel connections: Diode measurement
Measure diode forward bias using the front panel
To make a diode measurement using the front panel:
1. Make the connections as shown in the previous figures.
2. Press the FUNCTION key.
3. Select Diode.
4. Press the TRIGGER key for 2 seconds and verify that the instrument is set to Continuous
Measurement.
The measurements start displaying on the front panel.
Settings available for diode measurements
See Diode measure settings (on page 2-30) for settings that are available when you are making diode
measurements.