Section
2: General operation Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
2-136 DMM7510-901-01 Rev. B / May 2015
Current Waveform Quick Setup
The Current Waveform Quick Setup helps you set up Digitize Current measurements and displays
the results on the Graph screen.
You are prompted to set the sample rate, count, and signal amplitude when you select the Current
Waveform Quick Setup. The instrument is reset before these settings are applied. The instrument
also verifies whether the front or rear terminals are selected. If the front terminals are selected, signal
amplitude is limited to 3 A; if the rear terminal is selected, amplitude is limited to 10 A.
The sample rate defines the precise acquisition rate at which the digitizing measurements are made.
The count sets the number of measurements to digitize when a measurement is requested.
The signal amplitude allows the instrument to select a fixed range that is large enough to measure
your signal.
When the settings are complete, the Graph screen is displayed with the measure results.
Interval Measure Quick Setup
The Interval Measure Quick Setup helps you set up measurements that occur at precisely timed
intervals. The settings you make depend on whether you are using a measure function or a digitize
function.
When you run this Quick Setup for a measure function, the instrument:
Is reset
Prompts you to select a function
Prompts you for the interval time
Prompts you for the number of readings (count)
Sets up the trigger timer and trigger model
Initiates the trigger model
Displays measurements on the Home screen and stores them in defbuffer1
When you run this Quick Setup for a digitize function, the instrument:
Is reset
Prompts you for the interval (sample rate)
Prompts you for the number of readings (count)
Sets up a trigger model
Initiates the trigger model
Displays the measurements on the Home screen and stores them in defbuffer1

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals