7½ Digit Graphical Sampling Multimeter Reference Manual Section 6:
SCPI command reference
DMM7510-901-01 Rev. B / May 2015 6-35
Details
This command sets the scale factor (m) for an mx + b operation for the selected measurement
function.
The mx + b math operation lets you manipulate normal display readings (x) mathematically according
to the following calculation:
y = mx + b
Where:
y is the displayed result
m is a user-defined constant for the scale factor
x is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
b is the user-defined constant for the offset factor
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STAT ON
Set the math function for voltage measurements
to mx+b.
Set the scale factor for voltage measurements to
0.80.
Set the offset factor to 50.
Enable the math function.
Also see
Calculations that you can apply to measurements (on page 3-7)
:CALCulate[1]:<function>:MATH:FORMat (on page 6-31)
:CALCulate[1]:<function>:MATH:MBFactor (on page 6-33)
:CALCulate[1]:<function>:MATH:STATe (on page 6-37)