6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
6-106 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated
This command sets the simulated reference temperature of the thermocouple reference junction.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Celsius: 23
Kelvin: 296.15
Fahrenheit: 73.4
Usage
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated <tempValue>
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated DEFault
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated MINimum
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated MAXimum
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated?
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated? DEFault
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated? MINimum
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated? MAXimum
The function to which the setting applies; see Functions
• Celsius: 0 to 65
• Kelvin: 273 to 338
•
Functions
Details
This attribute applies to the temperature function when the transducer type is set to thermocouple and
the reference junction is set to simulated. It allows you to set the simulated temperature value.
Example
TEMP:TC:TYPE K
TEMP:UNIT C
Sets 30 degrees Celsius as the simulated
reference temperature for thermocouples.
Also see
[:SENSe[1]]:<function>:TCouple:TYPE (on page 6-107)
[:SENSe[1]]:<function>:TRANsducer (on page 6-112)
[:SENSe[1]]:<function>:UNIT (on page 6-113)
Temperature measurements (on page 2-120)