6: SCPI command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
6-106 DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated
This command sets the simulated reference temperature of the thermocouple reference junction.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Celsius: 23
Kelvin: 296.15
Fahrenheit: 73.4
Usage
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated <tempValue>
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated DEFault
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated MINimum
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated MAXimum
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated?
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated? DEFault
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated? MINimum
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated? MAXimum
<function>
The function to which the setting applies; see Functions
<tempValue>
The temperature:
• Celsius: 0 to 65
• Kelvin: 273 to 338
•
Fahrenheit: 32 to 149
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This attribute applies to the temperature function when the transducer type is set to thermocouple and
the reference junction is set to simulated. It allows you to set the simulated temperature value.
Example
TEMP:TRAN TC
TEMP:TC:TYPE K
TEMP:UNIT C
TEMP:TC:RJUN:SIM 30
Sets 30 degrees Celsius as the simulated
reference temperature for thermocouples.
Also see
[:SENSe[1]]:<function>:TCouple:TYPE (on page 6-107)
[:SENSe[1]]:<function>:TRANsducer (on page 6-112)
[:SENSe[1]]:<function>:UNIT (on page 6-113)
Temperature measurements (on page 2-120)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals