Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-19
Details
You cannot assign user-defined reading buffers the name defbuffer1 or defbuffer2.
If you create a reading buffer that has the same name as an existing user-defined buffer, the existing
buffer is overwritten by the new buffer. Any data in the existing buffer is lost.
When you create a reading buffer, it becomes the active buffer. If you create two reading buffers, the
last one you create becomes the active buffer.
The default fill mode of a user-defined buffer is once. You can change it to continuous.
Once the buffer style is selected, it cannot be changed.
Once you store the first reading in a compact buffer, you cannot change certain measurement
settings, including range, display digits, and units; you must clear the buffer first.
Not all remote commands are compatible with the compact, writable, and full writable buffer styles.
Check the Details section of the command descriptions before using them with any of these buffer
styles.
Writable readings are used to bring external data into the instrument. You cannot assign them to
collect data from the instrument.
You can change the buffer capacity for an existing buffer through the front panel or by using the
bufferVar.capacity command.
Example
capTest2 = buffer.make(200, buffer.STYLE_COMPACT)
Creates a 200-element compact reading buffer named
capTest2
.
Also see
bufferVar.capacity (on page 8-22)
bufferVar.fillmode (on page 8-28)
buffer.write.format() (on page 8-43)
buffer.write.reading() (on page 8-45)
Reading buffers (on page 3-13)
Remote buffer operation (on page 3-30)
Writable reading buffers (on page 3-34)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB