7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-23
Details
This command allows you to change or view how many readings a buffer can store. Changing the
size of a buffer will cause any existing data in the buffer to be lost.
The overall capacity of all buffers stored in the instrument cannot exceed 11,000,000 readings for
standard reading buffers and 27,500,000 for compact reading buffers. For more information about
buffer capacity, see Setting reading buffer capacity (on page 3-18
).
Example
reset()
testData = buffer.make(500)
capTest = buffer.make(300)
bufferCapacity = capTest.capacity
print(bufferCapacity)
print(testData.capacity)
testData.capacity = 600
print(testData.capacity)
print(defbuffer1.capacity)
Create two user-defined reading buffers: testData and
capTest.
Create a variable called bufferCapacity to hold the
capacity of the capTest buffer.
Print bufferCapacity.
Output:
300
Print the capacity of testData.
Output:
500
Changes the capacity of testData to 600.
Print the capacity of testData.
Output:
600
Print the capacity of the default buffer defbuffer1.
Output:
100000
Also see
buffer.delete() (on page 8-16)
buffer.make() (on page 8-18)
bufferVar.clear() (on page 8-24)
print() (on page 8-231)
printbuffer() (on page 8-232)
Reading buffers (on page 3-13)
Remote buffer operation (on page 3-30)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB