Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-75
Example
dmm.digitize.func = dmm.FUNC_DIGITIZE_VOLTAGE
dmm.digitize.range = 90
dmm.digitize.analogtrigger.mode = dmm.MODE_EDGE
dmm.digitize.analogtrigger.edge.level = 5
dmm.digitize.analogtrigger.edge.slope = dmm.SLOPE_FALLING
Set the function to digitize voltage.
Set the range to 90, which will select a range of 100 V.
Set the analog trigger mode to edge.
Set the level to sense 5 V.
Set the level to be detected on a falling edge.
Also see
Analog triggering overview (on page 3-64)
dmm.measure.analogtrigger.mode (on page 8-121)