Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-144 DMM7510-901-01 Rev. B / May 2015
Details
Use this command to store the active measure settings to a configuration index in a configuration list.
If the index parameter is not provided, the configuration index will append to the end of the list.
Configuration lists are not saved when the instrument is turned off. To save a configuration list, create
a configuration script to save instrument settings, including any defined configuration lists.
You cannot store both digitize and measure function settings in a configuration list that is used with a
trigger model.
Example
dmm.measure.configlist.store("MyConfigList")
Stores the active settings of the
instrument to the end of the configuration
list MyConfigList.
dmm.measure.configlist.store("MyConfigList", 5)
Stores the active settings of the
instrument to configuration index 5 in the
measure configuration list
MyConfigList
.
Also see
Configuration lists (on page 3-37)
dmm.measure.configlist.create() (on page 8-139)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals