EasyManua.ls Logo

Keithley DMM7510

Keithley DMM7510
1036 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
8: TSP command reference Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
8-182 DMM7510-901-01 Rev. B / May 2015
Minimum Maximum
DC voltage
-1000 1000
AC voltage
-700
700
DC current (rear terminals selected)
-10
10
DC current (front terminals selected)
-3
3
AC current (rear terminals selected)
-10
10
AC current (front terminals selected)
-3
3
Resistance
-1e+09 1e+09
4-wire resistance
-1e+09 1e+09
Diode
-10
10
Capacitance
-0.001
0.001
Temperature
-3310
3310
Continuity
-1000
1000
Frequency
-1e+06 1e+06
Period
-1 1
DC voltage ratio - Method set to result
-1E+12
1E+12
DC voltage ratio - Method set to parts
-1000
1000
Digitize voltage
-1000
1000
Digitize current (rear terminals selected)
-10
10
Digitize current (front terminals selected)
-3
3
If you have math, limits, or filter operations selected, you can set the relative offset value to include
the adjustments made by these operations. To include these operations, set
dmm.measure.rel.level to dmm.measure.read(). The adjustments from these operations are
not used if you use the dmm.measure.rel.acquire() function to set the relative offset level.
Example
dmm.measure.func = dmm.FUNC_DC_CURRENT
dmm.measure.rel.level = dmm.measure.read()
dmm.measure.rel.enable = dmm.ON
Set the measure function to DC current.
Set the relative offset level to be the reading
with any calculations included.
Enable the relative offset.
Also see
Relative offset (on page 3-4)
dmm.digitize.rel.level (on page 8-113)
dmm.measure.rel.acquire() (on page 8-179)
dmm.measure.rel.enable (on page 8-180)

Table of Contents

Other manuals for Keithley DMM7510

Related product manuals