Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 8:
TSP command reference
DMM7510-901-01 Rev. B / May 2015 8-203
Trigger events
Event description Event constant
No trigger event
trigger.EVENT_NONE
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFYN
A command interface trigger (bus trigger):
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
trigger.EVENT_COMMAND
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
(1 to 6)
trigger.EVENT_DIGION
Line edge detected on TSP-Link synchronization line N
(1 to 3)
trigger.EVENT_TSPLINKN
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)
trigger.EVENT_LANN
Trigger event blender N (1 to 2), which combines trigger
events
trigger.EVENT_BLENDERN
Trigger timer N (1 to 4) expired
trigger.EVENT_TIMERN
Analog trigger
trigger.EVENT_ANALOGTRIGGER
External in trigger
trigger.EVENT_EXTERNAL
Example
reset()
dmm.measure.func = dmm.FUNC_DC_CURRENT
dmm.trigger.measure.stimulus =
trigger.EVENT_DISPLAY
dmm.measure.count = 10
Reset the instrument.
Set the function to DC current.
Set the trigger stimulus to be the
front-panel TRIGGER key.
Set the count to 10.
Press the TRIGGER key to generate 10
readings that are stored in the active
reading buffer.
Also see
None