TriggerMXO 4 Series
175User Manual 1335.5337.02 ─ 07
Trigger out ← Set trigger out
Activates the outgoing pulse on the [Trigger Out] connector on the rear panel.
Remote command:
TRIGger:ACTions:OUT:STATe on page 571
Signal source ← Set trigger out
Defines when the trigger out signal is initiated: at the trigger point, when waiting for the
trigger, or when the post-trigger time is finished.
Remote command:
TRIGger:ACTions:OUT:SOURce on page 572
Polarity ← Set trigger out
Sets the polarity of the trigger out pulse, which is the direction of the first pulse edge.
Remote command:
TRIGger:ACTions:OUT:POLarity on page 572
Delay ← Set trigger out
Defines the delay of the first pulse edge to the trigger point. The minimum delay is
600 ns.
Remote command:
TRIGger:ACTions:OUT:DELay on page 572
Pulse length ← Set trigger out
Sets the length of the trigger out pulse.
Actions on trigger