TRIGger:EVENt<ev>:INTerval:WIDTh <Width>
Sets the time between two pulses for the interval trigger.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Width> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 5E-09
Default unit: s
Usage: Asynchronous command
Manual operation: See "Interv. width" on page 159
17.9.10 Slew rate trigger
TRIGger:EVENt<ev>:LEVel<n>:SLEW:LOWer..................................................................557
TRIGger:EVENt<ev>:LEVel<n>:SLEW:UPPer...................................................................557
TRIGger:EVENt<ev>:SLEW:DELTa..................................................................................558
TRIGger:EVENt<ev>:SLEW:RANGe................................................................................558
TRIGger:EVENt<ev>:SLEW:RATE...................................................................................559
TRIGger:EVENt<ev>:SLEW:SLOPe.................................................................................559
TRIGger:EVENt<ev>:LEVel<n>:SLEW:LOWer <Level>
Sets the lower voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: -10 to 10
Increment: 0.001
*RST: -0.1
Default unit: V
Usage: Asynchronous command
Manual operation: See "Lower level" on page 160
TRIGger:EVENt<ev>:LEVel<n>:SLEW:UPPer <Level>
Sets the upper voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event