Remote control commandsMXO 4 Series
878User Manual 1335.5337.02 ─ 07
17.19.3 STATus:QUEStionable registers
The commands of the STATus:QUEStionable subsystem control the status reporting
structures of the STATus:QUEStionable registers.
See also: Chapter 16.5.3.4, "STATus:QUEStionable register", on page 468.
The query of the [:EVENt] and [:ENABLe] commands, return a list of the affected
sources:
Table 17-8: Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:
[:ENABLe]
Bits Source values Description
For all bits
NONE
ALL
All available sources are activated
PPSupply PRobe<m>
Active probe
POVerload PRobe<m>
Active probe
COVerload CHAN<m>
Analog channels
WARNCHAN<m>
Warning overload channels
EXTTRIGGERIN
External analog signal connected to the external
trigger input
TRIGGEROUT
Trigger out signal
TEMPerature TEMPWARN
Temperature warning: indicates that the device is
getting hot and should be cooled.
TEMPERROR
Temperature error: the device is too hot and will
automatically shut down in a few seconds.
ADCState CPCHAN<m>
Positive clipping for analog channels
CNCHAN<m>
Negative clipping for analog channels
CPPRobe<m>
Positive clipping for probes
CNPRobe<m>
Negative clipping for probes
GOVerload WGENerator<m>
Available generators
NOALigndata HCHannel<m>
Horizontal channels
VCHannel<m>
Vertical channels
PROBe<m>
Probes channels
LPRObes<m>
Logic probes
WGENerator<m>
Available generators
LIMit MEAS<m>
Enabled measurements
MARGin MEAS<m>
Enabled measurements
IMPRecise MEAS<m>
Enabled measurements
Some of the status commands can be used interchangeably: