Remote control commandsMXO 4 Series
554User Manual 1335.5337.02 ─ 07
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time Width - Delta and for Width + Delta at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than Width - Delta or longer than Width +
Delta.
SHORter
Triggers if the signal crosses vertical limits before the specified
width time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
width time is reached.
*RST:
WITHin
Usage: Asynchronous command
Manual operation: See "Time condition" on page 156
TRIGger:EVENt<ev>:WINDow:WIDTh <Width>
Depending on the values of TRIGger:EVENt<ev>:WINDow:RANGe the width sets:
For the ranges Within and Outside, the width defines the center of a time range.
The range is defined by the limits "±Delta", see TRIGger:EVENt<ev>:WINDow:
DELTa.
For the ranges Shorter and Longer, it defines the maximum and minimum time
lapse, respectively.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Width> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 5E-09
Default unit: s
Usage: Asynchronous command
Manual operation: See "Width" on page 156
17.9.8 Timeout trigger
TRIGger:EVENt<ev>:TIMeout:RANGe............................................................................. 555