11.3 Digital voltmeter
The software-based voltmeter features AC RMS, DC and DC RMS voltage measure-
ments on a specific analog channel.
11.3.1 Using the digital voltmeter
To start a digital voltmeter measurement
1. Tap the "Menu" > "Apps" > "General" tab > "DVM", to open the digital voltmeter
measurement dialog.
2. Select the channel for your measurement.
3. Select a "Filter bandwidth" from the list.
4. Set a "Measurement time".
5. Tap on a measurement to enable it. Available are "DC", "DC RMS", "AC RMS".
The measurement results of all enabled measurements are displayed in a result
box.