Remote control commandsMXO 4 Series
624User Manual 1335.5337.02 ─ 07
17.12.5 Gate
GATE<g>:ENABle..........................................................................................................624
GATE<g>:GCOupling..................................................................................................... 624
GATE<g>:CURSor......................................................................................................... 625
GATE<g>:ZDIagram.......................................................................................................625
GATE<g>:MODE........................................................................................................... 625
GATE<g>:ABSolute:STARt............................................................................................. 626
GATE<g>:ABSolute:STOP..............................................................................................626
GATE<g>:RELative:STARt..............................................................................................626
GATE<g>:RELative:STOP.............................................................................................. 626
GATE<g>:SHOW........................................................................................................... 626
GATE<g>:ENABle <First>
Enables the gate.
Suffix:
<g>
.
1...2, index of the gate
Parameters:
<First> ON | OFF
Usage: Asynchronous command
Manual operation:
See " Add" on page 221
GATE<g>:GCOupling <CouplingMode>
The gate coupling mode selects how the gate area is defined.
Suffix:
<g>
.
1...2, index of the gate
Parameters:
<CouplingMode> MANual | CURSor | ZOOM | SPECtrum
MANual
Manually define the gate with a user-defined start and stop val-
ues.
CURSor
Cursor coupling is available if a cursor is defined. The gate area
is defined by the cursor lines of an active cursor measurement.
ZOOM
Zoom coupling is available if a zoom is defined. The gate area is
defined identically to the zoom area - if you change the zoom,
the gate changes as well.
SPECtrum
Spectrum coupling is available if a spectrum is enabled.
*RST:
MANual
Usage: Asynchronous command