EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Segmentation Settings

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquisition and waveform setupMXO 4 Series
115User Manual 1335.5337.02 ─ 07
Remote command:
ACQuire:AVERage? on page 516
Interpolation
Selects the interpolation method.
If the defined sample rate ("Sample rate") is higher than the ADC sample rate, interpo-
lation adds points between the captured samples of the waveform by various mathe-
matic methods. The selected interpolation method is also used for zooming.
"Linear"
Two adjacent ADC sample points are connected by a straight line, the
interpolated points are located on the line. You see a polygonal wave-
form similar to the real signal, and also the ADC sample points as
vertexes.
"sin (x)/x"
Two adjacent ADC sample points are connected by a sin(x)/x curve,
and also the adjoining sample points are considered by this curve.
The interpolated points are located on the resulting curve. This inter-
polation method is precise and shows the best signal curve.
"Sample/Hold"
The ADC sample points are displayed like a histogram. For each
sample interval, the voltage is taken from the sample point and con-
sidered as constant, and the intervals are connected with vertical
lines. Thus, you see the discrete values of the ADC - the measured
samples.
Remote command:
ACQuire:INTerpolate on page 517
6.2.3 Segmentation settings
Access: "Menu" > "Acquisition" > "Segmented" tab.
You can acquire a limited number of segments, or the maximum number.
Acquisition

Table of Contents

Related product manuals