Remote control commandsMXO 4 Series
517User Manual 1335.5337.02 ─ 07
ACQuire:CURRent?
Returns the current number of acquisitions that have been acquired.
Return values:
<CurrAcqCnt> Range: 0 to 18446744073709551615
Increment: 1
*RST: 0
Usage: Query only
Asynchronous command
ACQuire:INTerpolate <IntpolMd>
Selects the interpolation method.
Parameters:
<IntpolMd> SINX | LINear | SMHD
LINear
Linear interpolation between two adjacent sample points
SINX
Interpolation with a sin(x)/x function.
SMHD
Sample/Hold causes a histogram-like interpolation.
*RST: SINX
Usage: Asynchronous command
Manual operation: See "Interpolation" on page 115
ACQuire:POINts[:VALue] <RecordLength>
Sets the record length, if ACQuire:POINts:MODE is set to MANual.
Parameters:
<RecordLength> Range: 1000 to 200E+6
Increment: 2
*RST: 1000
Default unit: pts
Usage: Asynchronous command
Manual operation: See "Record length" on page 113
ACQuire:POINts:ARATe?
Returns the sample rate of the ADC, before waveform processing. The result is the
interleaved sample rate or the non-interleaved one, depending on the channel usage.
Acquisition and setup

Table of Contents