17.7.5 Display
Persistence........................................................................................................... 498
Signal.................................................................................................................... 499
Backlight................................................................................................................500
Clear results..........................................................................................................500
Save/Recall...........................................................................................................501
17.7.5.1 Persistence
DISPlay:PERSistence:INFinite.........................................................................................498
DISPlay:PERSistence:RESet.......................................................................................... 498
DISPlay:PERSistence:TIME............................................................................................498
DISPlay:PERSistence[:STATe].........................................................................................499
DISPlay:PERSistence:INFinite <State>
If infinite persistence is enabled, each new waveform point remains on the screen until
this option is disabled. Use infinite persistence to display rare events in the signal.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Infinite persistence" on page 90
DISPlay:PERSistence:RESet
Resets the display, removing persistent all waveform points.
Usage:
Setting only
Asynchronous command
Manual operation: See "Reset" on page 90
DISPlay:PERSistence:TIME <Time>
Sets a time factor that controls how long the waveforms points fade away from the dis-
play. Thus, the MXO 4 emulates the persistence of analog phosphor screens.
Parameters:
<Time> Range: 0.05 to 50
Increment: 0.05
*RST: 0.05
Default unit: s
Usage: Asynchronous command
Manual operation: See "Time" on page 90